SUGGESTED MEDICAL GAS MASTER ALARM POINTS

 

FOR LEVEL 1 FACILITIES

 

 

CRYOGENIC BULK GAS UNITS WITH CRYOGENIC RESERVE

    1.  LIQUID LEVEL LOW

    2.  RESERVE IN USE

    3.  MAIN LINE PRESSURE HIGH

    4.  MAIN LINE PRESSURE LOW

    5.  RESERVE LOW

    6.  CHANGEOVER

    7.  RESERVE PRESSURE LOW

 

CRYOGENIC BULK GAS UNITS WITH CYLINDER RESERVE

    1.  LIQUID LEVEL LOW

    2.  RESERVE IN USE

    3.  MAIN LINE PRESSURE HIGH

    4.  MAIN LINE PRESSURE LOW

    5.  RESERVE PRESSURE LOW (Not required if reserve is cylinders

          with check valve on each.)

    6.  CHANGEOVER

 

MANIFOLDS WITH RESERVE

    1.  CYLINDER BANK CHANGEOVER

    2.  RESERVE IN USE

    3.  RESERVE PRESSURE LOW

    4.  MAIN LINE PRESSURE HIGH

    5.  MAIN LINE PRESSURE LOW

 

MANIFOLDS WITHOUT RESERVE

    1.  CYLINDER BANK CHANGEOVER

    2.  MAIN LINE PRESSURE HIGH

    3.  MAIN LINE PRESSURE LOW

 

MEDICAL AIR COMPRESSORS (ALL TYPES)

    1.  RESERVE COMPRESSOR RUNNING (see note 1)

    2.  MAIN LINE PRESSURE HIGH

    3.  MAIN LINE PRESSURE LOW

    4.  CARBON MONOXIDE HIGH (see note 1)

    5.  DEW POINT HIGH

 

ADDITIONAL SIGNALS REQUIRED DEPENDING ON COMPRESSOR

DESIGN:

 

For Liquid Ring Compressors:

    1.  RECEIVER FLOODED (see note 1)

    2.  HIGH WATER SEPARATOR (see note 1)

 

For Reciprocating (Piston Type) Compressors:

    1.  HIGH TEMPERATURE (see note 1)

    2.  RECEIVER FLOODED (see notes 1 and 2)

 

MEDICAL/SURGICAL VACUUM PUMPS

    1.  RESERVE VACUUM PUMP RUNNING

    2.  MAIN LINE VACUUM LOW

 

WASTE ANESTHETIC GAS VACUUM PUMPS

    1.  RESERVE VACUUM PUMP RUNNING (see note 3)

    2.  MAIN LINE EVACUATION LOW (see note 3)

 

NOTES:

    1.  Per NFPA-99, 1999 edition, these alarm points can be

          monitored with a single alarm point on each of the Master

          Alarm Panels with a label such as “Source Equipment

          Abnormal”.

    2.  For Reciprocating Air Compressors with Water Cooled Heads

          and/or Water Cooled Aftercoolers.

    3.  These alarm points are not required per NFPA-99, 1999 Edition

 

 

Medical Gas.